OP03 Procedures op03:

HRC Higher-Level Sequences

Use Radioactive Calibration Source

Description:

This sequence contains the steps necessary to perform a measurement using the HRC's internal radioactive calibration source. The HRC detector housing door must be positioned so that the source illuminates the detectors and the cal source is moved to its exposed position. After the measurement is finished the cal source is stowed and the door moved back to the open position.

Cautions/Restrictions:

The HRC shutters must be in their "Home" position before this sequence is run (see HRC-C-001).

Flow Diagram

Procedure