OP03 Procedures | op03: |
This sequence should be used to bring the HRC instrument into a "stand-by" mode independent of which detector, HRC-I or HRC-S, is to be made active.
If the CTU is currently operating in format 1 only the first set of serial digital housekeeping values in a major frame will be valid. In order to be sure of having completely valid HRC serial digital telemetry this sequence should be followed by the appropriate sequence to set the telemetry format.
Delta-Time | Command Mnemonic | Description | Cmd Field Mnemonic | Data Parameter | Telemetry Verifier | Expected Value | Notes |
---|---|---|---|---|---|---|---|
00:00:01.000 | 205VAON | +5V Low Voltage Power Supply On | 2PS5AON | ON | hlp3e
After +5V is on the serial digital commands to HRC will be affective |
||
00:00:01.000 | 2SPHVOF | HRC-S HVPS Off | 2SPONST | OFF | sd0100
Precautionary: at +5V turn on the HVPS should be default to "off" |
||
00:00:01.000 | 2SPTTHV | Set HRC-S top MCP HV step to 0 | 2SPTTHV2 | 0 | 2SPTPAST | 0 | sd0200 |
00:00:01.000 | 2SPTBHV | Set HRC-S bottom MCP HV step to 0 | 2SPTBHV2 | 0 | 2SPBPAST | 0 | sd0300 |
00:00:01.000 | 2SPCLEN | Enable HRC-S HVPS current limit | 2SPCLST | ENAB | sd0401 | ||
00:00:01.000 | 2S1HVOF | Shield PMT#1 HVPS off | 2S1ONST | OFF | sd0500
Precautionary: at +5V turn on the HVPS should be default to "off" |
||
00:00:01.000 | 2S1STHV | Set shield PMT#1 HV step | 2S1STHV2 | 0 | 2S1HVST | 0 | sd0600 |
00:00:01.000 | 2IMHVOF | HRC-I HVPS Off | 2IMONST | OFF | sd0900
Precautionary: at +5V turn on the HVPS should be default to "off" |
||
00:00:01.000 | 2IMTTHV | Set HRC-I top MCP HV step to 0 | 2IMTTHV2 | 0 | 2IMTPAST | 0 | sd0a00 |
00:00:01.000 | 2IMTBHV | Set HRC-I bottom MCP HV step to 0 | 2IMTBHV2 | 0 | 2IMBPAST | 0 | sd0b00 |
00:00:01.000 | 2IMCLEN | Enable HRC-I HVPS current limit | 2IMCLST | ENAB | sd0c01 | ||
00:00:01.000 | 2S2HVOF | Shield PMT#2 HVPS off | 2S2ONST | OFF | sd0d00
Precautionary: at +5V turn on the HVPS should be default to "off" |
||
00:00:01.000 | 2S2STHV | Set shield PMT#2 HV step to 0 | 2S2STHV2 | 0 | 2S2HVST | 0 | sd0e00 |
00:00:01.000 | 2MDRVADI | Disable motor drive electronics | 2MDRVAST | DISA | sd6800 | ||
00:00:01.000 | 2MCMRASL | Motion control mode reset | 2MCMDARS | RSET | sd6f00 | ||
00:00:01.000 | 2ALMTADS | Deselect all motors | 2ALMTAST | SEL | sd7000 | ||
00:00:01.000 | 215PCAON | +/-15V low voltage power supply on | 215PCAST | ON | sd1001
Dual sided supply; this command turns on both +15V and -15V |
||
00:00:02.000 | 224PCAON | +24V low voltage power supply on | 224PCAST | ON | sd1201 | ||
00:01:30.000 | 2PRA2SL | Shield PMT#2 select to use A-side pre-amplifiers | 2ELEASS | PMT2 | hlp3d | ||
00:00:01.000 | 2PRB1SL | Shield PMT#1 select to use B-side pre-amplifiers | 2ELEBSS | PMT1 | hlp7c | ||
00:00:01.000 | 2NXILASL | Select timing for HRC next-in-line mode telemetry | 2OBNLASL | NEXT | sd1a01 | ||
00:00:01.000 | 2FIFOAOF | Reset event data fifo | 2FIFOAVR | RSET | sd4500 | ||
00:00:01.000 | 2FIFOAON | Enable event data FIFO | 2FIFOAVR | ENAB | sd4501 |
$Log: power-up.html,v $
Revision 1.8 1998/06/08 17:48:50 juda
corrected spelling error "Rest" to "Reset"
Revision 1.7 1998/05/18 15:53:42 juda
corrected telemetry verifier mnemonics for MCP HV Step commands (2IMTPST -> 2IMTPAST, 2IMBPST -> 2IMBPAST, 2SPTPST -> 2SPTPAST and 2SPBPST -> 2SPBPAST)
Revision 1.6 1998/05/08 19:31:46 juda
Added "SCS #69" to sequence name
Revision 1.5 1998/05/07 14:53:46 juda
Converted "side A" reference to "CEA A" - added File block to version table
Revision 1.4 1998/05/06 21:43:43 juda
Added PMT preamp selection and default next-in-line timing, FIFO reset and enable commands
Revision 1.3 1998/05/05 12:59:31 juda
revised based on new EGSE script received from J. Gomes on 1998 May 4 - removed many of the serial digital configuration commands (these will go with the instrument specific sequences) - removed 215NCAON
Revision 1.2 1998/04/30 18:27:57 juda
Moved motor "safing" commands 2MDRVADI (sd6800), 2MCMRASL (sd6f00), and 2ALMTADS (sd7000) to before +/-15V and +24V turn-on (recommendation of Jack Gomes during an informal discussion)