The quantum efficiency of the ACIS FI CCDs decreases at high energies and far from the read-out as a result of the CTI. This effect is readily visible in the temperature maps of hot clusters (e.g. Coma). Here we investigate the energy- and CHIPY-dependence of this effect using calibration observations of the Coma cluster and SNR G21.5-0.9 at the focal plane temperature -110C.
At 6 keV, the results derived here are in agreement with the lab measurement by C. Grant in both the amplitude of the effect and its CHIPY dependence.
The effect is prominent between 3 and 10 keV and increases with energy.
There are two independent studies of the CTI effect on the CCD QE using the calibration source and both lab models (Catherine Grant at MIT) and in-flight data (Leisa Townsley at PSU).
Both these studies consistently indicate a decrease of the QE at the energy of the Mn Ka line (6 kev) but not at the lower-energy lines (0.5-1.5 keV). Also, the QE at 5.9keV seems to reach a ``saturation'' value of at large CHIPY compared to CHIPY=0 (i.e., near the readout).
Unfortunately, these results cannot be directly used for the calibration of the position-dependent CCD QE because the QE is measured essentially at only one energy. The energy dependence of the CTI effect on the QE can be established using observations of the same X-ray source in different regions of the CCD.
The SNR G21.5-0.9 was observed in Nov-Dec of 1999 in I0, I1, I2, I3, and S2 chips (Fig. 2.1). The average source CHIPY for each observation is presented in Table 1. Assuming that the HRMA energy-dependent vignetting is well-calibrated we can use the I0 pointing as a reference and determine the QE in the middle of I1, I2, and S2, and at large CHIPY's for I3.
The angular diameter of the SNR is 1 arcmin, which is much smaller than the angular scale at which the CTI effect is expected to vary.
Coma was observed in Nov-Dec of 1999 in a series of pointings in which the source position was fixed with respect to the HRMA and ACIS was moved along the Z-axis by approximately 3 arcmin. With this setup, the same region of the Coma cluster was observed in two pointings in S2 at the same off-axis angle, but at different CHIPY's. This opens a possibility to derive the relative QE without assuming that the vignetting is perfectly known. Unfortunately, Z motions are parallel to the CHIPX direction in the ACIS-I CCDs, and therefore the QE measurement cannot be performed with them.
We defined three rectangular regions which are within the S2 field of view in both pointings (Fig. 2 and Table 2). Using region #1, we can derive the relative change in the CCD QE between CHIPY=175 and 464, with r2 -- from 390 to 681, and with r3 -- from 600 to 890. Combining the data for all regions, we can derive the QE change essentially across the entire CHIPY>400 region.
|
Since we want to compare the observed spectra of G21.5-0.9 and the Coma
cluster in different chip positions, we first need to correct for the
CTI-caused gain variations across the CCDs. I used
acis_process_events
with the acisD1999-09-16gainN0003.fits
gain table to recompute photon energies and PI channels.
I then extracted the source spectra in PI channels in the same regions in sky coordinates. The background spectra were produced from the blank sky data provided by M. Markevitch. RMFs were generated from the -110C FEFs (the FEFs were modified to correct the problem with negative Gaussian components), and ARFs were calculated using the latest HRMA area calibration files and the spatially uniform CCD QE files distributed with CIAO. The QE curves are individually calibrated for each CCD and presumably are accurate at the readout.
For perfectly calibrated CCD QE, HRMA vignetting, and RMFs, these spectra should produce identical fits in XSPEC.
I have chosen the I0 observation as a reference because it is the closest to the readout. The source spectrum was well-fit with the broken power law plus absorption model (Fig 3) with , the lower energy photon index , the higher-energy photon index , and the break energy keV. The deviations of the data from the model are no more than 10% at keV. I then used the same model (including normalization) for I1,I2,I3, and S2.
The result for the I3 pointing in shown in Fig. 4 and the ratio is shown separately in Fig. 5. There is a clear deficit of the data relative to the model at keV. The data around 1.5 keV may indicate some residual gain variations of order 1-2%, but they cannot explain the effect at high energies.
Note that the analytic model in Fig. 3 and 4 was convolved with the RMFs generated for the appropriate CCD regions; therefore, the effect of the varying energy resolution was corrected at least in the first order.
The I1, I2, and S2 spectra of G21.5-0.9 (all taken in the middle of the chip) show the same effect as that in Fig 4-5 -- the ratio between the observed spectrum and the model assuming uniform QE decreases approximately linearly with from at keV to at keV, and the trend seems to continue to higher energies reaching at 8-9 keV.
If this is interpreted as the QE decrease due to CTI we conclude that
The last two points perfectly agree with what is found in the lab model at -110C (see Catherine Grant's web page).
For the Coma cluster, I fit the obsid# 556 data with the Raymond-Smith plus absorption model and then apply the same model to the obsid#1086 data which is at the same off-axis angle but at higher CHIPY. In this case, the ``reference'' spectra in obsid 556 are already affected by the QE decrease and we are measuring the relative change of the QE between the respective chipy positions.
For the region 3 (CHIPY change 175 to 464), the effect is very similar to that in G21.5-0.9 (Fig. 6). The region 2 data indicate, albeit marginally, the 10% change of QE at keV between CHIPY=390 and 690 (Fig 7). Finally, the region 1 data indicate that there is no detectable change in the QE between CHIPY=605 and 890.
Thus, the Coma cluster results fully support the conclusions conclusion from the G21.5-0.9 data (page ).
Here we assume that the effects seen in the G21.5-0.9 and Coma data are entirely due to the CTI-induced decrease in the CCD QE. Figure 9 shows the CHIPY dependence of the QE at keV derived from the lab measurements in MIT, PSU in-flight calibration source data averaged over all I3 nodes, and the in-flight astronomical source data described here.
The conclusions can be summarized as follows: